GENERAL SPECIFICATIONS
• Dimensions with pads:
- 30 cm (11.81 in) length.
- 21.5 cm (8.46 in) depth.
- 28.0 cm (11.02 in) height.
• Weight:
- Device – 4.90 Kg (10.80 lbs).
- Battery – 1.15 Kg (2,53 lbs).
- External pads – 0.85 Kg (1,87 lbs).
- Complete equipment (except for NIBP) – 6.90 Kg (15,21 lbs).
• Electrical:
- AC: 100 to 265 VAC, 50/60 Hz (automatic selection).
- DC external: 10 to 16 VDC.
• Removable battery:
- Type: NiMH, 14.4 VDC 4 A/h.
- Duration: Battery with full charge - 3 hours in monitor mode, without printer, or a minimum of 140 shocks at 360 joules or a minimum of 200 shocks at 200 joules.
• Memory:
- Type: Nand Flash.
- Capacity: 2 Mbytes.
- Patients stored >150 patients.
- Storage: 15 seconds of ECG when in shock, physiological alarm and panel events.
ENVIRONMENTAL SPECIFICATIONS
• Temperature:
- Operational: 0 to 50 ºC.
- Storage: 0 to 70 ºC.
• Humidity:
- Operational: 10 to 95% RH, without condensation.
- Storage: 10 to 100% RH, without condensation.
DEFIBRILLATOR
• Waveform: Biphasic truncated exponential. Waveform parameters adjusted in terms of patient's impedance.
• Adult / external defibrillation:
- Scales: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 40, 50, 80, 100, 150, 200, 250, 300 and 360 Joules. Maximum power limited to 50 J with internal or children's pads.
•Controls: On / off button, charge, shock, synchronism.
• Power selection: Therapy button on the front panel.
• Charge control: Button on the front panel, button on the external pads.
• Shock control: Button on the front panel, buttons on the external pads.
• Synchronized control: SYNC button on the front panel.
• Charge indicators:
- Sound signal of equipment being charged.
- Sound signal of completed charge.
- LED on external pads and charge level indicated on the display.
• Maximum charging time: Network and battery < 6s.
• Electrode size:
- Adult: 10,3 cm x 8,5 cm. Contact area: 81,9 cm2
- Children: 4,5 cm x 4,0 cm. Contact area: 18 cm2
• Cardioversion: < 60ms.
• Pads (options):
- Adult and child external (included).
- Adult and child internal: (optional).
- Multifunctional for pacemaker, monitoring and defibrillation (optional).
- Multifunctional extension (optional).
AED MODULE (OPTIONAL)
• Functional characteristics: voice command, visual indications, CPR instructions, USB 2.0 for synchronism with PC. Multilanguage, Sudden Death Prevention Technology (SDP).
• USB: USB 2.0 for transfer of the electrocardiogram stored in AED mode to a compatible PC.
• SoftDEA: Software for viewing the data transferred to the PC.
EXTERNAL PACEMAKER (OPTIONAL)
• Modes: Demand or fixed.
• Amplitude: From 5 mA to 200 mA (resolution of 5 mA), accuracy 10%.
• Pulse width: 20 ms (+/- 1ms).
• Frequency: From 30 ppm to 180 ppm (increments of 5 ppm), accuracy ± 2%.
• Refractory period:
- 340 ms (from 30 to 80 ppm).
- 240 ms (from 90 to 180 ppm). |
NIBP (OPTIONAL)
• Operating principle: Oscillometric.
• Automatic measurement mode: 1, 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes.
• Manual: One measurement.
• Measurement interval:
• Adult/pediatric:
- Systolic: 30 - 255 mmHg.
- Mean: 20 - 235 mmHg.
- Diastolic: 15 - 110 mmHg.
- Pulse: 40 - 240 BPM.
• Neonatal:
- Systolic: 30 - 135 mmHg.
- Mean: 20 - 125 mmHg.
- Diastolic: 15 - 110 mmHg.
- Pulse: 40 - 240 BPM.
• Overpressure limit by software:
- Adult: 290 ± mmHg.
- Neonatal: 145 mmHg max.
• Overpressure protection by hardware:
- Adult: 300 ± mmHg.
- Neonatal: 150 ± mmHg.
• Resolution: 1 mmHg.
DISPLAY
• Battery level indicator: Yes.
• Size: 99.36 mm x 132.48 mm.
• Diagonal: 6,5”.
• Type: Color LCD TFT.
• Resolution: 640 x 480 pixels (VGA).
• Scan speed: 12.5; 25 and 50 mm/s.
ECG
• Inputs:
- 3 or 5 lead ECG cable.
- External pads
- Multifunctional pads.
• Range: 15 to 300 bpm.
• Precision: +/- 1 bpm from 15 to 300 bpm.
• Rejection in common mode: Greater than 90 dB, measured according to AAMI standards for heart monitors (EC 13).
• Sensitivity: 5, 10, 15, 20, 30 and 40 mm/mV.
• AC line filter: 60 Hz or 50 Hz.
• ECG response frequency:
- Diagnostic mode - (0.05 - 100 Hz).
- Monitor Mode - (1 - 40 Hz).
• Patient insulation: Defibrillation proof
- ECG: CF Type.
- SpO2: CF Type.
• Loose Electrode: Identified and shown with low level alarm.
• Defibrillator discharge: < 5 seconds.
SpO2 (OPTIONAL)
• SpO2 Range: 0 to 100 %.
• Pulse Range: 30 to 250 bpm.
• SpO2 Precision:
+/- 2 % from 70 to 100%..
+/- 2 % from 70 to 100%.
• Pulse Precision: +/- 2 bpm.
• Scan speed: 12.5; 25 and 50 mm/s.
PRINTER (OPTIONAL)
• Type: Thermal.
• Weight: 0.4 Kg.
• Speed: 12.5; 25 or 50 mm/s with precision of ±5%.
• Paper size: 50 mm (width) x 30 m (maximum length).
STANDARDS:
- EN 60601-1:1990 (A1:1993, A2:1995, A13:1996)
- NBR IEC 60601-1:1994 + emenda 1997
- EN 60601-1-2:2007
- NBR IEC 60601-1-2:2006
- EN 60601-2-4:2003
- NBR IEC 60601-2-4:2005
- EN 60601-2-25:1995 (A1:1999)
- EN 60601-2-27:2006
- NBR IEC 60601-2-27:1997
- EN 60601-2-49:2001
- NBR IEC 60601-2-49:2003
|