CONFIGURATIONS:
See the table.
CHARACTERISTICS:
• Fast initialization: 10 seconds
• Alarms:
- Priorities: 2 types (low and high)
- Notification: Audible and visual
• Intensity: 10 levels (multiple tones)
• Alarm silence: adjustable in 30, 60,120 and 180 sec
• Fully customizable screen exhibition
• 72 hours of tabular and waveform
trend
• Autoset for alarms
• Operation modes: Neonatal, pediatric and adult
• NIBP automatic measurement
TREND:
• Types: graphic and tabular
• Memory: 72h (non-volatile)
• Time between data: 40 sec and 40 samples
• Graphics: one waveform per vital sign
TECHNICAL CHARACTERISTICS:
NIBP:
• Principle of operation: oscillometric
• Automatic measurement mode: 1, 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes
• Manual: one measurement
• Stat mode: consecutive
measurements in 5 minutes
• Measurement interval:
- Adult/pediatric:
- Systolic: 30 - 255 mmHg
- Mean: 20 - 235 mmHg
- Diastolic: 15 - 110 mmHg
- Pulse: 40 - 240 BPM
- Neonatal:
- Systolic: 30 - 135 mmHg
- Mean: 20 - 125 mmHg
- Diastolic: 15 - 110 mmHg
- Pulse: 40 - 240 BPM
• Software overpressure limit:
- Adult: 290 mmHg max
- Neonatal: 145 mmHg max
• Hardware overpressure protection:
- Adult: 300 ± mmHg
- Neonatal:150 ± mmHg
• Resolution: 1 mmHg
SPO2:
• Measurement range: 0 - 100%
- Accuracy:
- ± 2% from 70 - 100%
- ± 3 % de 50 - 69%
• Pulse:
- Range: 30 - 250 BPM
- Accuracy: ± 2 BPM from 30 - 250
BPM
• Scan speed: 12, 5, 25 and 50 mm/s
ECG:
• Sensitivity: 5, 10, 15, 20, 30 e 40
mm/mV
• Scan speed: 12.5, 25 or 50 mm/s
• Derivations: DI, DII, DIII, aVR, aVL,
aVF, C
• ECG cable: 3 or 5 leads
• Filter: 35 Hz and 60 Hz
• Measurement range: 15 to 250 BPM
• Accuracy: ± 1 BPM from 30 to 250
BPM
• Loose electrode indication
• Recognition of pacemaker pulse
• Defibrillator discharge protection: < 5 sec according to IEC601-2-27
RESPIRATION:
• Technical: Transtoraxic impedance
• Measurement range: 3 to 150
resp/min
• Accuracy: ± 3 resp/min
• Sensitivity: 1, 2, 3, 4, 5, and 6
• Electrodes: RA - LA
INVASIVE BLOOD PRESSURE:
• Filter: 50Hz and 60Hz
• Measurement interval: form -
99 mmHg to 310 mmHg
• Zeroing interval: ± 70 mmHg
• Accuracy: ± 1%, ± 1 digit, which is
bigger
• Transducer: 5μV/V/mmHg,
disposable or reusable
CAPNOGRAPHY:
• Measurement range: 0 - 99 mmHg
• Accuracy:
± 3 mmHg from 0 - 40 mmHg
± 8% from 41 - 76 mmHg
± 10% from 77 - 99 mmHg
• Calibration method: two points
• Start:
- 10 seconds for CO2 Curve start
- Less than 3 minutes for complete functioning
• Consumption: 0,8 W
• Compensation: BTPS, N2O, O2 |
TEMPERATURE:
• Measurement range: 15ºC - 45ºC (59ºF - 113ºF)
• Resolution: ± 0,1 ºC (0,18ºF)
• Compatible with series YSI 400 for
temperature sensors
NORMAS:
• NBR IEC 60601-1:1994/1997
• NBR IEC 60601-1-2:1997
• NBR IEC 60601-2-27:1997
• NBR IEC 60601-2-30:1997
• NBR IEC 60601-2-49:2003
• ANSI/AAMI EC13:2002
• ANSI/AAMI SP 10
SPECIFICATIONS:
• Weight: 4.2 Kg (excluding accessories)
• Dimensions:
- Height: 231 mm.
- Width: 300 mm.
- Depth: 190 mm.
COLOR SCREEN:
• Type: LCD TFT COLOR (Active
Matrix)
• Backlight: 2 CCFL Lamps (Cold
Cathode Fluorescent Lamp)
• Size: 8,4" or 10,4"
• Traces: 6 simultaneous traces (maximum)
• Screen view:
- Intelligent interface that automatically
adjusts itself for the number of present
parameters, always offering the best
visualization.
RECORDER (optional):
• Type: Thermal
• Weight: 0.4 Kg
• Paper width: 50 mm
• Max. Length: 30 meters
BATTERY:
• AC 85 - 265 VAC, 50/60 Hz
• DC 10 - 16 VDC
POWER SUPPLY:
• AC 85 - 265 VAC, 50/60 Hz
• DC 10 - 16 VDC
OPERATING CONDITIONS:
• Temperature:
- Operating: 0 - 50 ºC
- Storage: 0 - 70 ºC
• Humidity:
- Operating: 10 - 95% RH, without
condensation
- Storage: 10 - 100% RH, without
condensation
• Electromagnetic Compatibility:
- Irradiated and conducted, CISPR11
ACCESSORIES:
BASIC:
• 01 power cable (3 pins) - code 555-0
(B)
• 01 auxiliary grounding cable and
potential equalizing - code 549-5 (B)
• 01 user manual - code 18805
ECG:
• 01 5-lead ECG cable - code 18376
SpO2:
• 01 oximetry sensor 3044 - code
1255-6 (D)
• 01 oximetry extension - code 18627
(H)
TEMPERATURE:
• 01 adult surface temperature
sensor - code 18384 (A)
NIBP:
• 01 adult cuff - code 18562 (F)
• 01 cuff extension - code 12432 (G)
IBP:
• 01 Medex transducer MX960
• 01 interconnect cable MX961Z14
• 01 transducer support MX261
• 01 Medex monitoring line kit
MX9604A
CAPNOGRAPHY:
• 01 low humidity monitoring line
• 01 adult/pediatric sample line “T” |