Automated External Defibrillator

*Please, note: "Isis" is also marketed as "Eazy" in selected markets.

DESCRIPTION

The Automated External Defibrillator (AED) Isis/Eazy, using its Neural Network Technology is capable of assessing, through sophisticated sensors, the patient's condition, considering the clinical variables and automatically applying the most adequate shock therapy. That allows an individual with basic training to assist a victim with cardiac

arrest, facilitating and enhancing the possibilities of salvation.

Acts in case of cardiac arrest (ventricular fibrillation and ventricular tachycardia), by voice direction, diagnosis and by automatic and safe application of shock treatment upon activation of a single button.

CHARACTERISTICS

  • Operation with just one button.
  • Artificial Intelligence: accurate diagnosis of the patient’s conditions, indicating or not application of shock and hindering accidental use.
  • Minimum of 200 shocks (200 Joules, full load, battery under good conditions).
  • Instructions by means of voice and visual indicators on the LCD screen.
  • Horizontal or vertical use.
  • Record of events for further analysis.
  • PC connection by USB.
  • Biphasic Shock.
  • Self-diagnosis of functions and battery.
  • Connection, download and data management software by PC.
  • Easy access to pads for use and replacement.

ADVANCED TECHNOLOGIES

Operation With Just One Button

Biphasic Shock

Artificial Intelligence

Just place the adhesive pads and press the single front button and follow the instructions by means of voice and visual indicators.

 

The two-phase wave form technology acts on ventricular fibrillation conditions and ventricular tachycardia reversion processes more efficiently, because it allows for the use of lower power, resulting in smaller myocardial lesions after treatment.

Artificial Intelligence: accurate diagnosis of the patient’s conditions, indicating or not application of shock and hindering accidental use.

VIDEO TUTORIAL

TECHNICAL INFORMATION

GENERAL SPECIFICATIONS

  • Dimensions:
  • 22.0 cm/8.66 in (W).
  • 13.0 cm/5.11 in (D).
  • 29.0 cm/11.41 in (H).
  • Weight:
  • Device - 2.90 kg.
  • Internal rechargeable battery:
  • Type: Li-ion, 14.4 Vdc 4.0 A/h.
  • Duration: 10 hours in cardiac rhythm recognition mode (full load battery) or at least 200 shocks in 200 Joules (battery under good conditions).
  • Battery full load time (fully unloaded): 5 hours.
  • Battery loader source:
  • Power supply 100 – 240V/50-60Hz.
  • Consumption (max.): Power supply 1 A.
  • Output: 24 Vdc, 1.5 A.
  • Battery storage:
  • Battery storage for long periods under temperatures above 35ºC will reduce the battery’s capacity and abbreviate its useful life.
  • Preset defibrillation scales:
  • Adult: 1st shock 150 J, subsequent shocks 200 J.
  • Child: 50 J.
  • Internal memory storage:
  • 100 events or 2 hours of ECG record.
  • Classification:
  • Internal Energized Equipment.
  • CF Type.

 

  • Mode of operation:

Continuous Operation.

  • Maximum time since beginning of rhythm analysis until promptness for discharge:
  • 20 s.
  • Maximum time since beginning of the operation until promptness for discharge at the maximum energy:
  • 25 s.

 

ENVIRONMENTAL SPECIFICATIONS

  • Temperature:
  • Operational: 0 to 50ºC/32 to 122 F.
  • Storage: 0 to 70ºC/32 to 158 F.
  • Humidity:
  • Operational: 10 to 95% RH, without condensation.
  • Storage: 10 to 95% RH, without condensation.

 

DEFIBRILLATOR

  • Wave form:
  • Biphasic Truncated Exponential. Wave form parameters set according to the patient’s impedance.
  • Shock application:
  • By multifunctional adhesive pads.
  • Commands
  • Front Panel button (on/off).
  • Defibrillation scales:
  • Adult: 150 and 200 J.
  • Child: 50 J.
  • Adult/Child Selection:
  • Automatic by the type of pads.
  • Charge command:
  • Automatic after identifying arrhythmias that should receive shock.
  • Shock command:
  • Front Panel button, when blinking.
  • Maximum charging time:
  • 50 Joules: < 2 seconds.
  • 150 Joules: < 4 seconds.
  • 200 Joules: < 6 seconds.
  • Pads size:
  • Adult = area: 82 cm2/12.71 in2.
  • Child = area: 30 cm2/4.65 in2.
  • Maximum output voltage:
  • 1500 V.
  • Maximum output current:

60 A (25 ohms).

 

STANDARDS

  • NBR IEC 60601-1
  • NBR IEC 60601-1-2
  • NBR IEC 60601-1-6
  • NBR IEC 60601-1-8
  • NBR IEC 60601-2-4

or equivalent IECs.